Humidity test chambers usually employ one of three dehumidification techniques to achieve lower humidity conditions. These methods may be used in temperature-only applications, as well, when risk of condensation is high due to massive test loads and/or very fast temperature cycling rates.
1. Refrigeration coil: Nearly every humidity chamber should have refrigeration to ensure humidity control. To start with, relative humidity varies with temperature, so a stable temperature is required. Most humidity chambers have a separate dehumidification coil to attract moisture. It is set-up to stay above freezing to avoid becoming covered with ice. The main cooling coil is also kept above freezing, as well.
2. Nitrogen purge: Nitrogen gas is very dry. By slowly bleeding nitrogen gas into the chamber, moisture-laden air is purged out of the chamber.
3. Dry-air purge: Because nitrogen isn't always readily available, a dry-air system is more common. External compressed air is fed into a desiccant dryer system before being used to purge the chamber air. (A desiccant is a substance that easily absorbs moisture, like the silica gel crystal packets you find packaged with some products.) The dry-air purge system doesn't need much maintenance. The system uses two desiccant “towers”. One is being used to dry the compressed air while the other tower is dried out (and the moisture vented to the room). You will often hear these systems “pop” while switching between towers. The desiccant can be used for years before replacement desiccant is needed.
De-humidification (or how to lower humidity)
免責聲明
- 凡本網注明“來源:化工儀器網”的所有作品,均為浙江興旺寶明通網絡有限公司-化工儀器網合法擁有版權或有權使用的作品,未經本網授權不得轉載、摘編或利用其它方式使用上述作品。已經本網授權使用作品的,應在授權范圍內使用,并注明“來源:化工儀器網”。違反上述聲明者,本網將追究其相關法律責任。
- 本網轉載并注明自其他來源(非化工儀器網)的作品,目的在于傳遞更多信息,并不代表本網贊同其觀點和對其真實性負責,不承擔此類作品侵權行為的直接責任及連帶責任。其他媒體、網站或個人從本網轉載時,必須保留本網注明的作品第一來源,并自負版權等法律責任。
- 如涉及作品內容、版權等問題,請在作品發(fā)表之日起一周內與本網聯(lián)系,否則視為放棄相關權利。